3. Time-out actions
The
5(7FRQWUROYLVF
VDIHW\FRQWUROmagnetic stirrer
by IKA Werke offers internal monitoring by
labworldsoft
®
or the connected computer as an additional safety feature. If the control software
or the computer fails (crash/power failure etc.) this can switch off the heating and stirring
function or switch to a safe temperature and speed. You can activate the watchdog, specify the
time slot and set the action to be triggered in the event of a fault in the shaker’s parameters pop-
up window.
The watchdog time-out must be at least 20 times greater than the active sample rate of the
measurement.
The safety monitoring functions are only available once the target speed control path has been
activated.
1RWH
Please see the separate operating instructions for more details of the magnetic stirrer
properties.